To Clear an Alarm
1. Correct any flow blockage.
2. Press the POWER ON/OFF button to restart the Pump.
Turning OFF the Universal Pump Probe
Press and hold the POWER ON/OFF button for three seconds.
•
The red LED flashes once per second.
•
The pump shuts down.
•
The audible alarm sounds.
Battery Warning
A Battery Warning indicates that a nominal 15 minutes of operation
remain before device batteries are completely depleted.
NOTE: Duration of remaining Pump Probe operation during Battery
Warning depends on:
•
Ambient temperatures
(The battery warning is likely to be shorter in colder
temperatures).
•
When the Pump Probe goes into Battery Warning:
•
Yellow and green LEDs alternately flash every other second.
•
The Pump continues to operate until the device is turned OFF
or battery shutdown occurs.
Battery Shutdown
When the batteries can no longer operate the device, the device goes
into Battery Shutdown mode:
•
Yellow LED flashes
•
Audible alarm sounds
•
Pump shuts down.