To Clear An Alarm; Calibration Check; Diffusion Instrument - MSA Orion Manual De Operaciones

Ocultar thumbs Ver también para Orion:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Operating Manual
5. Press the CHANGE/RESET button to reset the alarm and
restart the pump.
During operation, a pump alarm may occur when the:
• Flow system is blocked
• Pump is inoperative
• Sample lines are attached or removed.

To Clear an Alarm

1. Correct any flow blockage.
2. Press the CHANGE/RESET button.
• The Pump will now restart.
NOTE:
When the instrument is in a gas alarm, the pump
alarm may not display until gas alarm is cleared.

Calibration Check

The calibration check is simple and should only take about one
minute. Perform this calibration check before each day's use.
1. Turn ON the Orion
Multigas Detector in clean,
fresh air.
2. Verify that readings
indicate no gas is present.

Diffusion Instrument

If your Orion Multigas Detector is
NOT equipped with an optional
built-in sampling pump:
1. Attach calibration cap to
the Orion Multigas Detector,
orienting the inlet fitting
toward the display
(FIGURE 3-4).
2. Attach regulator (supplied with
calibration kit) to the cylinder.
Chapter 3, Using the Orion Multigas Detector
Figure 3-4. Instrument with
Calibration Cap Installed
3-7

Publicidad

Tabla de contenido
loading

Tabla de contenido