Advanced operations
5.2.3
Software MM-485
5.3
PROFIBUS DP/V0 Output
5.3.1
Connection
66
This is a very simple software that controls the communication line RTS. It allows viewing
and saving in a file (txt format) the data sent by one or several instruments MultiMeter 44.
PC requirements: Windows 98 or superior.
Installation instructions
1. Insert the CD in the CD-ROM unit of the PC.
2. Copy the file MM485.exe directly in the PC and execute it. It is required to select the
communication type, RS485 or RS232, the language and the number of the serial
port.
Utilization
1. When start up the application, it will ask for the file name where the data will be
stored. If this file already exists, there will be a message informing that the file
content will be substituted.
2. If the series channel (COM) has never been selected or it is not correct, the
application will ask to introduce it.
3. Select the communication type with the PC, RS 485.
4. Select the language.
5. Press the key "Instruments" and introduce the serial numbers of the connected
instruments. If the serial numbers are introduced wrongly, the application will not
recognized the sent data.
Lines with the sent data by the connected multimeters appear on the main window of the
application.
to stop storing data in the selected file.
•
CLOSE FILE
to erase the data appearing on the display. (It does not affect the
•
CLEAR DISPLAY
data already stored).
to quit the application.
•
QUIT
Note: The data sending interval should be selected in every connected instrument, see "Data
Logger, Capture interval" in Instrument configuration.
Always disconnect power to the instrument when making any electrical connections.
Relays: Limit the current up to 3A by using one fuse. Do not connect cables with voltage higher
than 24 V.
D A N G E R
D A N G E R