7.1 Troubleshooting the controller

7.2 Troubleshooting the analyzer

7.2.1 LED status

LED status
green LED
orange LED
LED flashing

7.2.2 Error messages

Instrument
Error displayed
reaction
Warms up and
Temperature <
goes into the
0°C/32°F
service state
Instrument goes
Analyzer is too
into the service
cold.
state
16
Troubleshooting
If entries are only implemented with a delay or are not accepted for a short time, the delay
may be caused by a busy data network. Refer to the troubleshooting section in the sc1500
User Manual.
If, in normal operation, problems occur that are apparently caused by the controller, restart
the system.
After a software update, a system expansion or after an interruption in the power supply, it
may be necessary to set the system parameters again.
Note all the values that are changed or entered so all the necessary data can be used to
configure the parameters again.
1. Save all important data.
2. Isolate the supply of power and wait for 5 seconds.
3. Reapply power to the controller.
4. Check all relevant settings.
5. If the problems still occurs, contact Technical Support.
If the complete analysis instrument is not working, check whether the humidity sensor has
tripped. Fix the damage, dry the humidity sensor and restart the system.
If the problems still occurs, contact Technical Support.
Table 5 LED Status and definitions
red LED
Cause
instrument was
below 4 °C (39 °F)
on power up
Instrument interior
has been below
4 °C (39 °F) for
more than
5 minutes
Definition
No errors of warnings
Warning
No communication with controller
Solution
Check whether instrument is frozen
(Cleaning solution/sample/reagent/
standards). If necessary use
pre-warmed reagents. Thaw electrode,
delete error. Instrument will then
continue to warm up and start
Close instrument, check heating
Error
Reset error
Reset error
manually
Test/Maintenance>
Reset error
Reset error
manually
Test/Maintenance>
Reset error