4.9 Show calibration information

1. Select SENSOR SETUP > [select sensor] > CALIBRATION > CAL DATA.
2. Select an option.
Option
LAST CAL
DATE
SLOPE
OFFSET

4.10 Set to factory calibration

To set the sensor back to the factory calibration, select SENSOR SETUP > [select sensor] >
CALIBRATION > SET CAL DEFLT.
22 English
Description
Shows the date of the last calibration.
Note: The production date of the digital gateway shows until the 2-point calibration
procedure is done.
Shows the slope value from the last calibration (or the value entered
manually).
Note: The slope value is 1.00 when the sensor is set to the factory calibration.
Shows the offset value from the last calibration (or the value entered
manually).
Note: The offset value is 0.0 µg/L PAH when the sensor is set to the factory calibration.