instruments. Do not use instruments and cables unless they
are designed to fit properly. Consult the high frequency
electrosurgical cable's user manual for additional
instructions.
Recommended HF Electrosurgical Cables Such As:
50-2111
Reusable Monopolar Cable
40-2110
Disposable Monopolar Cable
Electrosurgical generators
Use a compatible electrosurgical generator for monopolar
surgery. Do not exceed 7000 volts peak-to-peak. Consult
the generator user manual for additional instructions.
Instructions for Use
Attach one end of a compatible sterile high frequency
electrosurgical cable to the laparoscopic manual instrument.
Approximate
Size
Color
Small
Purple
< 6.5
Small
Purple
< 6.5
Medium
Green
7-8
Medium
Green
7-8
Large
Blue
> 8.5
Large
Blue
> 8.5
(A) Flush Port
(B) Rotation Knob
(C) Electrosurgical Post
(D) Position Adjustment Button
(E) Handle Grip
(F) Trigger
(G) Insert Release Button
(H) Shaft
(I) Insert
IFU‐LCN‐ERG100\Rev C
Working
Glove Size
Length
33cm
45cm
33cm
45cm
33cm
45cm
Repositioning
1.
Retract the position adjustment button (D).
2.
Rotate handle to desired position.
3.
Release the position adjustment button (D).
Assembly
1.
With handle in the pistol-grip position, depress the
insert release button (G), move trigger (F) fully
forward into the eject position.
2.
While holding the trigger (F) forward, let go of
insert release button (G) and slide insert (I) into
distal end of shaft.
3.
Allow insert (I) to push trigger (F) back while
aligning up insert pegs with slots of shaft (H).
4.
Once insert release button (G) engages (pops up),
pull trigger (F) closed until inserts clicks into place.