determinations to validate the reagent blank value. If the reagent blank value is not accurate, the analyzer will
not give accurate results during normal operation.
Note: Do not use ZERO CAL with Hach reagents. The reagent blank value of Hach reagents (molybdate) is
measured during controlled conditions at the factory and is very accurate. To enter the reagent blank value for
Hach reagents, use the OVERRIDE CALIBRATION option (refer to
3. The measurement status is displayed. Select YES to interrupt the current measurement cycle and
start the calibration immediately. Select NO to wait until the current measurement cycle has
completed before the calibration starts.
4. Complete the instructions on the display.
Calibration
on page 19).