Segment display
Graphical program operation display:
Value
1.1
1.2
2
3
⇒ There is only either one heating ramp (1.1) or one cooling ramp (1.2) per firing segment - never both!
6.3.2.
Example of a firing program to illustrate the firing segment
Example of a simple firing program to illustrate the firing segments and their values:
Firing program
Firing
Firing segment values
segment
Program delay or program start delay
0
(see Section 6.6.3.)
Heating ramp (1.1)
1
Target temperature (2)
Soak time (3)
Heating ramp (1.1)
2
Soak temperature (2)
Soak time (3)
Cooling ramp (1.2)
3
Target temperature (2)
Soak time (3)
Cooling ramp (1.2)
4
Target temperature (2)
Soak time (3)
Graphical program
operation display
Meaning
Heating ramp
Cooling ramp
Soak Temperature (target temperature)
Soak time
Graphical program operation display
Description of the values
The firing program starts with a delay.
(Factory-set to 00 hours 00 min.)
Heating at 250°C/h
Heating up to 500°C
Hold for 0 min
Heating at 0°C/h
Hold at 500°C
Hold for 40 min
Cooling at 40°C/h
Cool to 20°C
Hold for 0 min
Setting the value to "END" ends the firing program.
-
-