Troubleshooting - HoMedics THERA-P BPW-200 Manual Del Usuario

Ocultar thumbs Ver también para THERA-P BPW-200:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Cet appareil numérique respecte les limites de bruits
radioeléctriques applicables aux appareils numériques de
Classe B prescrites dans la norme sur le matériel brouilleur:
"Appareils Numériques", ICES-003 édictée par le ministre
des communications.
33

TROUBLESHOOTING

If any abnormality arises during use, please check the following
points:
POSSIBLE
SYMPTOMS
CAUSES
Unit does not
Batteries have
turn on when the
run down.
START/STOP
button is pushed.
Battery polarities
have been posi-
tioned incorrectly.
EE measurement
Cuff has been
error symbol shown
placed incorrectly.
on display or the
blood pressure
Did you talk or move
value is displayed
during
excessively low
measurement?
(or high).
Shaking of the wrist
with the cuff on.
E1 error symbol
Air circuit
shown on display
abnormality.
E2 error symbol
Inflation pressure
shown on display
exceeding 300 mmHg.
E3 error symbol
Data error.
shown on display
Note: If the unit still does not work, contact HoMedics Consumer
Relations. Under no circumstance should you disassemble
or attempt to repair the unit by yourself. Contact informa-
tion for HoMedics Consumer Relations Department can be
found on the warranty page (See page 36).
CORRECTION
Replace them
with two new AAA
alkaline batteries.
Re-insert the
batteries in the
correct positions.
Wrap the cuff prop-
erly so that it is posi-
tioned correctly.
Measure again.
Keep wrist
steady during
measurement.
Measure again.
Switch the unit off,
then measure again.
Remove and reload
the batteries.
34

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido