Cleaning Of Accessories; Cleaning Device - Gima CORSIA Manual De Istrucciones

Ocultar thumbs Ver también para CORSIA:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Proceed as follows to disinfect the accessories:
1.
Turn the upper part of the nebulizer in an anti-clockwise direction;
2.
Disconnect the internal pisper at the base of the nebulizer using the fingers;
3.
After using the appliance, disassemble the nebulizer and clean all parts in warm water, rinse carefully and remove excess water using a soft
cloth and leave to dry in a clean place.
4.
To disinfect the accessories (except the tube) use denatured alcohol or hypochlorite solutions easily obtainable from chemists. Before using
after cleaning, rinse the accessories in warm water until all traces of the disinfectant have been eliminated, dry and store in a dry and dust
free place.
After each treatment clean thoroughly each component of the nebulizer (except air tube) removing medication residual and possible impurities.
Clean all parts in warm water. Rinse thoroughly making sure that all deposits are washed away and let dry
The nebulizer HI-FLO, the mouthpiece and the nosepiece must be disinfected by boiling in water (for max. 10 minutes).
Attention: Never use a cleaning brush or put sharp objects into the jet holes as this will damage the nebulizer
Use a soft dry cloth with not – abrasive and not – solvent detergents. The device's plug must be removed from the wall socket before proceeding
with any cleaning procedures.
The aerosol CORSIA is intended for use in the electromagnetic environment specified below. The customers or the user of the aerosol CORSIA
Emissions Test
Irradiated / Conducted emissions CISPR11
Irradiated / Conducted emissions CISPR11
Harmonic emissions IEC/EN 61000-3-2
Voltage fluctuations / flicker emissions
IEC/EN 61000-3-3
The aerosol CORSIA is intended for use in the electromagnetic environment specified below.
The customers or the user of the aerosol CORSIA should assure that it's used in such an environment.
Immunity Test
Electrostatic discharge
(ESD)
IEC/EN 61000-4-2
Electrical fast transient
/ burst
IEC/EN 61000-4-4

CLEANING OF ACCESSORIES

DO NOT BOIL OR PUT IN AUTOCLAVE MASK AND AIR TUBE

CLEANING DEVICE

Guidance and manufacturer's declaration – Electromagnetic Emissions
should assure that it's used in such an environment.
Compliance
Group 1
Class [B]
Class [A]
Complies
Guidance and manufacturer's declaration – Immunity Emissions
Level indicated by the
IEC 60601-1-2
± 6kV on contact
± 8kV in air
± 2kV power supply lines
± 1kV for input / output lines
1-
Air Tube
2-
Nebulizer Tank
3-
Nebulization Nozzle
4-
Nebulizer Top
5-
Mouthpiece
6-
Adult Mask
7-
Pediatric Mask
8-
Nosepiece
Electromagnetic environment - guidance
The aerosol CORSIA only used RF energy only for its internal
functioning. Therefore its RF emissions are very low and are not
cause interference in proximity of any Electronic appliances.
The aerosol CORSIA can be used in all environments, including
domestic and those connected directly to the public mains distribution
that supplies power to environments used for domestic scopes or
environments feeds to you from batteries.
Compliance Level
Electromagnetic environments - guidance
The device doesn't
Floors should be wood, or ceramic tile. If floors
change its state
are covered with synthetic material, the relative
humidity should be at least 30%.
The device doesn't
Mains power quality should be that of a typical
change its state
commercial environment or hospital
14

Publicidad

Tabla de contenido
loading

Tabla de contenido