2 Disinfect the prismatic tube of the CEREC Bluecam with the HLD set.
Not approved in the USA
Approved in the USA
6
Operating instructions for CEREC Bluecam and CEREC Omnicam Care and HLD (high-level disinfection)
2
Disinfect the prismatic tube of the CEREC
Bluecam with the HLD set.
NOTICE
Fragment: no water in camera
Make sure that no water gets into the camera when rinsing the prismatic
tube.
The complete process for high-level disinfection (HLD) is as follows:
Cleaning the camera
➢ Cleaning process after using the CEREC Bluecam Camera: Wipe the
camera with the appropriate prismatic tube so that the superficial
impurities cannot harden and stick to the surface of the prismatic
tube. Use a lint-free cloth as listed below which has first been soaked
in one of the cleaning products.
Agent approved for USA/not approved for USA
Alpro
Merz
Dürr
Kerr Corporation
Rinsing the exterior of the prismatic tube
➢ Rinse the exterior of the prismatic tube.
Sirona Dental Systems GmbH
● Minuten Spray classic
● Minuten Wipes
● Plasti Sept
● Plasti Sept Wipes
● Pursept-A
● FD 312
● CaviCide
● Cavi Wipes
D3492.201.02.02.02 01.2016
62 82 839 D3492