Operating instructions for CEREC Bluecam and CEREC Omnicam Care and HLD (high-level disinfection)
Sterilizing
Refitting the mirror sleeve
62 82 839 D3492
D3492.201.02.02.02 01.2016
Sterilizing
CAUTION
If the CEREC Omnicam accidentally falls down, check to make sure that
the camera window and coated sapphire glass are not damaged. If the
CEREC Omnicam has been damaged, it must no longer be used on
patients.
The CEREC Omnicam must be recalibrated.
NOTICE
Not sterilizable!
Do not under any circumstances sterilize the CEREC Omnicam or the
video cable!
The mirror sleeve can be sterilized with hot air (180°C, 30 min) (not in the
autoclave!).
NOTICE
RKI note
For markets where the RKI* guidelines are to be observed
The mirror sleeve is classified as a "semicritical medical device A"
according to RKI guidelines and therefore does not have to be
autoclavable.
*RKI=Robert Koch Institute, Berlin (Germany).
Refitting the Mirror Sleeve
NOTICE
Risk of damaging the camera window or the coated sapphire glass.
➢ The mirror sleeve must not come into contact with the camera
window.
➢ Push the mirror sleeve straight toward the camera body; do not tilt it.
➢ Carefully refit the mirror sleeve until it locks in place.
3 Caring for the CEREC Omnicam and mirror sleeve