Laryngeal Mirror; Care Instructions; General Information; Cleaning And Disinfection - Riester uni I econom Instrucciones Para El Uso

Ocultar thumbs Ver también para uni I econom:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 68
the depressor.
Take the tongue blade holder and place it on the front of the bent arm illuminator.
Take standard wooden or plastic depressors and push them into the opening provided in
the tongue blade holder.
Please note that the temperature can exceed 41 °C (105,8°F) during intended use!

3.9 Laryngeal mirror

3.9.1. Startup and function
1. Place the laryngeal mirror on the lamp holder.
The laryngeal mirrors can only be used in combination with the bent arm illuminator.
This ensures optimal illumination. Take one of the 2 laryngeal mirrors and attach it to the
front of the bent arm illuminator.
Please note that the temperature can exceed 41 °C (105,8°F) during intended use!

4. Care instructions

4.1 General information

Cleaning and disinfecting the medical devices serves to protect the patient, user, third par-
ties, and maintain the integrity of the devices.
The product design and materials used make it impossible to define an upper limit on max.
feasible treatment cycles. The service life of medical devices is determined by their function
and careful handling.
Before return for repair, defective products must have undergone the prescribed reproces-
sing procedure.
If a reusable device shows signs of material deterioration, it should no longer be used and
should be disposed of according to the procedures described in the disposal/warranty sec-
tions.

4.2 Cleaning and disinfection

To avoid possible cross-contamination, diagnostic instruments and their handles must be
cleaned and disinfected regularly.
The diagnostic instruments together with their handles can be cleaned on the outside using
a damp cloth (if necessary, moistened with alcohol) until they are visibly clean. Wipe with
disinfectant (e.g. disinfectant Bacillol AF from Bode Chemie GmbH (time 30s)) only accor-
ding to the instructions of use of the respective manufacturer of disinfectant. Only disin-
fectants with proven effectiveness according to national guidelines should be used. After
disinfecting, wipe the instruments with a damp cloth to remove potential residue.
Please make sure that the cloth is moistened but NOT wet, so that no moisture penetrates
the openings in the diagnostic instrument or its handle.
• Make sure that glass and lenses are only cleaned with a dry and clean cloth.
The diagnostic instruments are not sterile devices; they cannot be sterilized
Never place the instrument heads and handles in liquids!
Make sure that no liquids penetrate the housing interior!

4.3 Processing of reusable ear specula

Cleaning: manual
Required equipment: mildly alkaline cleaner (e.g. neodisher Mediclean, Dr. Weigert 404333
has been validated) 15°C-50°C, cleaning brush (Interlock 09098 and 09050 have been va-
lidated), tap water/running water 20±2°C of at least drinking water quality, tub/basin for
36

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Uni ii economUni iii econom

Tabla de contenido