Riester uni I econom Instrucciones Para El Uso página 41

Ocultar thumbs Ver también para uni I econom:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 68
Guidelines and manufacturer's declaration - electromagnetic emissions
The diagnostic instruments uni® I, II, III / econom® with accessories are intended for
use in an electromagnetic environment as specified below. The customer or user of the
diagnostic instruments uni® I, II, III / econom® with accessories should ensure that
these are used in such an environment.
Immunity testing
Electrostatic
discharge (ESD)
IEC 61000-4-2
Fast transient
electrical distur-
bances / bursts
IEC 61000-4-4
Impulse voltage
IEC 61000-4-5
IEC 61000-4-11
Voltage dips,
short interrupti-
ons and voltage
variations accor-
ding to
IEC 61000-1-11
Magnetic field
with energy-ef-
ficient rated
frequencies
IEC 61000-4-8
Note: U
is the AC source. Mains voltage before the application of the test level.
T
IEC 60601
test level
Con: ±8 kV
Air: ±2, 4, 8,
15 kV
5/50 ns, 100
kHz, ±2 kV
±0.5 kV voltage
Phase-to-pha-
se conductor
±2 kV voltage
Outer conduc-
tor to earth
<0% UT 0.5
period at
0, 45, 90, 135,
180, 225,
270 and 315
degrees
0% UT 1 peri-
od and 70% UT
25/30 periods
Single-phase:
at 0 degrees
(50/60 Hz)
30A/m
50/60 Hz
Electromagnetic environment
Compliance
- guidance
Con: ±8 kV
Floors should be wood, concrete
or ceramic tile.
Air: ±2, 4, 8,
If the floor is covered with
15 kV
synthetic material, the relative
humidity must be at least 30%.
Not applicable
Not applicable
Not applica-
bled
30A/m
50/60 Hz
The quality of the supply voltage
should be that of a typical com-
mercial or hospital environment.
The quality of the supply voltage
should be that of a typical com-
mercial or hospital environment.
The quality of the supply
voltage should be that of a
typical commercial or hospital
environment.
Mains frequency magnetic
fields should be at a level
characteristic of a typical lo-
cation in a typical commercial
hospital environment.
41

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Uni ii economUni iii econom

Tabla de contenido