cleaning agent, lint-free cloths (Braun Wipes Eco 19726 have been validated).
1. The cleaning solution is prepared according to the manufacturer's instructions for the
cleaning agent (neodisher Mediclean 0.5% has been validated).
2. Completely immerse the medical devices in the cleaning solution.
3. Make sure that all surfaces are completely wetted with cleaning solution.
4. Carry out all subsequent steps below the liquid level to prevent the contaminated liquid
from splashing.
5. Brush the hard-to-reach areas of the immersed ear specula with a soft brush during the
exposure time. Pay attention to the critical, hard-to-reach places where a visual assess-
ment of the cleaning effect is not possible.
6. The total exposure time in the cleaning solution is 10 minutes (10 minutes has been va-
lidated).
7. Remove the medical devices from the cleaning solution.
8. Rinse the medical devices under running tap water (at least drinking water quality) for
at least 1 minute (1 minute has been validated) to completely remove any supernatant
or residual cleaning solution. Check that the device is clean; if soiling is visible, repeat
the above steps.
9. Dry with a lint-free cloth.
Disinfection: manual
Required equipment: Disinfectant (e.g. CIDEX OPA, Johnson & Johnson 20391 has been va-
lidated), demineralised water (demineralised water free of facultative pathogenic microor-
ganisms according to the KRINKO/BfArM recommendation) 20±2°C, sterile, lint-free cloths.
1. Prepare the disinfectant solution according to the manufacturer's instructions (CIDEX
OPA is a ready-to-use solution; the concentration must be checked using test strips, see
manufacturer's instructions) (CIDEX OPA has been validated).
2. Completely immerse the ear specula in the disinfectant solution.
3. Exposure time of the disinfectant solution according to the manufacturer's instructions
for high-level disinfection (CIDEX OPA for 12 minutes has been validated).
4. Remove the ear specula from the disinfectant solution and place them in a tub/basin
containing demineralised water for 1 minute (1 minute has been validated).
5. Repeat the step twice with fresh demineralised water.
6. Place the ear specula on a clean, dry cloth and allow to dry.
Further information for the user:
For information on cleaning and disinfection, refer to the current DIN EN ISO 17664 stan-
dard.
The homepage of RKI Guideline – KRINKO/BfArM also regularly provides information about
developments regarding cleaning and disinfection for the reprocessing of medical devices.
CAUTION!
The article is not approved for machine reprocessing and sterilization. This can lead to ir-
reparable damage!

5. Technical specifications

Vacuum 2.7 V (2.5 V is specified)
Xenon 2.5 V
Xenon 3.5 V
Lifespan:
Socket handle:
Power supply:
Operation
Degree of protection
Classification
Ambient temperature:
Transport:
Air pressure:

6. Spare parts and accessories

OTOSCOPE
No. 11504
Cover glass for otoscope
No. 10448
Swivel lens for otoscope
Specula for otoscope
300 mA
700 mA
700 mA
Approx. 15 hours
Optional 230 V or 120 V version
Continuous operation
Protection class II
Application part type B
0° to + 40° 30% to 70% non-condensing
-10° to + 55° 10% to 95% RH
800 hPa - 1100 hPa
37