Restoring factory settings (rESEt)
3. Select a filtering level.
FIL
Weight display
0
Sensitive
1
Medium
2
Slow
NOTE:
• If "0" is set, then if you are weighing patients who are not very steady
on their feet, it is possible that despite the"Hold" function being acti-
vated, no weight value will be permanently displayed.
• The greatest deviation between the weight value displayed and the
actual value occurs with setting "2".
4. Confirm your selection.
You exit the menu automatically.
You can restore the factory settings for the functions below.
Function
Autohold (AHOLd")
Beep (PrESS)
Beep (HOLd)
Beep (MEASr)
Filtering (FIL)
Autoclear (ACLr)
Pre-tare (Pt)
Display backlighting brightness
Display backlighting duration
BMI/BFR
Physical activity level (PAL)
Age in years
Age in months
Unit for height
Wireless module (SYS)
Autosend (ASEnd)
Autoprint (APrt)
Beeps for measurement
Language
Patient instructions
Announcement of measured results
Volume
Announcement "Take your printed ticket" (Print)
Ultra
NOTE:
The wireless module is switched off when factory settings are
restored. Information about existing wireless groups is retained.
Wireless groups do not have to be set up again.
1. Select the item "rESEt" from the menu.
2. Confirm your selection.
You exit the menu automatically.
3. Switch off the scale.
Factory settings are restored and are available when the scale is switched
back on.
Hold
Slow
Medium
Fast
Factory setting
Off
On
On
On
0
On
0 kg
50 %
Permanent
BMI
1.0
18
0
cm
Off
Off
Off
On
Variant-dependent
On
Variant-dependent
50 %
Off
Active
Operation • 89