Battery Handle Type Aa With Rheotronic, 2; Inserting The Batteries; Removing The Batteries - Riester EliteVue Instrucciones Para El Uso

Ocultar thumbs Ver también para EliteVue:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 56
Function:
Instrument head with battery handle type C and ri-accu® USB is ready for use while charging.
This allows for it to be used in a patient examination while charging.
CAUTION!
If a patient examination is done while charging (battery handle type C with USB charging techno-
logy, art.no. 10704), only the Riester medical power supply (art.no. 10709) may be used, since this
ME system was tested in accordance with standard IEC 60601-1: 2005 (Third Edition) + CORR.1:
2006 + CORR.2: 2007 + A1: 2012 and is medically approved.
This ri-accu® USB has a charge status indicator.
LED is green: battery is fully charged.
LED flashes green: battery is being charged.
LED is orange: battery is too weak and must be charged.
- Specifications:
18650 Li-ion battery, 3.6 V, 2600 mAh, 9.62 Wh
Ambient temperature:
Relative humidity:
Transport and storage temperature:
Relative humidity:
Air pressure:
Operating environment:
The ri-accu® USB is used exclusively by professionals in clinics and medical practices.
2.7 Battery handle type AA with rheotronic®, 2.5 V
To operate these battery handles, you need 2 standard alkaline AA batteries (IEC standard LR6)
2.8 Battery handle type AA with rheotronic®, 3.5 V (for ri-charger® L) To operate this battery
handle you need:
− 1 Riester rechargeable battery, 3.5 V (art.no.10690 ri-accu® L)
− 1 ri-charger® L charger (art.no. 10705, art.no. 10706)
3. Operation (inserting and removing batteries and rechargeable batteries)
CAUTION!
Use only the combinations described under 2.3 to 2.8!

3.1 Inserting the batteries:

Battery handles (2.3 and 2.7) type C and AA with rheotronic® 2.5 V:
− Unscrew the battery handle cover on the lower part of the handle in an anticlockwise direction.
− Insert the standard alkaline batteries designated for this battery handle with the plus side in
the direction of the handle top into the battery handle.
− Firmly screw the battery handle cover back on.

3.2 Removing the batteries:

− Battery handles (2.3 and 2.7) type C and AA with rheotronic® 2.5 V:
− Unscrew the battery handle cover on the lower part of the battery handle in an anticlockwise
direction.
− Remove the batteries from the battery handle by holding the opening of the battery handle
slightly downwards and shaking it slightly if necessary.
− Firmly screw the battery handle cover back on.
CAUTION!
For all factory-installed or separately supplied batteries, the red safety foil on the plus side must
be removed before startup!
CAUTION!
Only applies to battery handle (2.5) type C with rheotronic® 3.5 V for charging at a 230 V or 120
V socket:
When using the new ri-accu® L, art.no. 10692, ensure that the spring of the battery handle cover
is not insulated. When using the old ri-accu® L, art.no. 10692, the spring must be insulated
(short circuit hazard).
neuer ri-accu
L
®
26
0 ° t + 40 °C
30% to 70% non-condensing
-10 ° to +55 °C
10% to 95% non-condensing
800 hPa - 1100 hPa
alter ri-accu
L
®

Publicidad

Tabla de contenido
loading

Tabla de contenido