Error Codes; Corrective Action - smartheart 01-517 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

ERROR CODES

Code
err 0
no pulse or detect pulses not enough.
err 1
Leakage in Cuff Pressure/ Inflation too
low.
err 2
Can't get systolic pressure.
err 3
Can't get diastolic pressure.
err 4
Can't figure out the pressure or it outsides
the legal range.
err 5
Deflation is too fast.
err 6
Deflation is too slow.
err 7
Pressure fault.
err 8
Pressure fault.
err 9
Cuff pressure fault.
22 • English
Meaning

Corrective Action

Take off heavy clothes and retry again.
The arm cuff is not fastened properly.
Re-apply the cuff, and take a measurement
again.
Rest a while, relax and retry again.
Rest a while, relax and retry again.
Rest a while, relax and retry again.
The arm cuff is not fastened properly.
Re-apply the cuff, and take a measurement
again.
The arm cuff is not fastened properly.
Re-apply the cuff, and take a measurement
again.
Rest a while, relax and retry again.
Rest a while, relax and retry again.
The arm cuff is not fastened properly.
Re-apply the cuff, and take a measurement
again.