ERROR CODES

Code
E0
Pressure system is unstable before
measurement.
E1 / E2
Failed to detect systolic/ diastolic
pressure.
E3/ E4
Pneumatic system leakage or cuff is too
loose/ too tight during inflation.
E5
The cuff pressure is above 300mmHg.
More Than 3 minutes with cuff pressure
E6
above 15 mmHg.
E7
EEPROM accessing error.
E8
Pressure Sensor parameter error.
ER A
Device Parameter checking error.
Meaning
Corrective Action
Review the 'Taking a Measurement'
portion of the manual, wait a minimum of
5 minutes and attempt a new
measurement.
Review the Applying Cuff portion of this
manual; re-apply the cuff, wait a minimum
of 5 minutes and attempt a new
measurement.
Check that the cuff is applied properly,
wait a minimum of 5 minutes and attempt
a new measurement. If the monitor still
does not perform or performs abnormally,
contact Customer Care at 866-326-1313.
ENGLISH • 27