Observation
The bottle goes up with the
piercing head
No antifoam being injected
26 English
Explanation
Stop the analysis by pressing the emergency STOP button.
Disconnect instrument from the power supply and remove the bottle. Check
that the piercing spike is in place and not unscrewed. Release emergency
STOP button by turning it clockwise and restart the instrument.
Open the front door to stop the process. Remove the bottle and check that the
piercing spike is in place and is not unscrewed. Close the front door and press
the Abort button on the screen.
Go to the antifoam section of the maintenance wizard and prime the pumps
until antifoam can be seen being regularly injected into the antifoam priming
beaker.