Option
Display diagnostic
results
Skip comments view
after analysis
Package size check
Warning messages
Normalization
Column selection
When all details have been entered, select Ok to confirm.
7.1.3.2 Other parameters
1. Time and Date - Set the system date and time parameters and the display format. Select Ok to
confirm.
2. System information - Display the system information about the currently configured boards.
Select Exit to exit the screen.
7.1.4 Security and user management
7.1.4.1 Security management
When the instrument is started for the first time, security is enabled. Refer to
is highly recommended that each user be entered into the system and given appropriate access
rights as soon as possible to avoid any unauthorized access.
1. Configure parameters related to confidentiality. On completion, select Ok to confirm.
Option
Access rights
enabled
Auto logoff
Audit trail
Description
Use this option to troubleshoot measurement problems. If this box is checked, then
instead of the standard measurement results screen being displayed at the end of
the measurement process, more detailed measurement values are displayed
instead.
If checked, the comments section on the measurement results screen is not
displayed.
If checked, the instrument will verify that the package size is the same as that
defined in the height parameter for the package being measured. If different a
warning message will be displayed. The height detection resolution is 1 mm.
If checked, any warning messages will be displayed during the package
measurement process. If left unchecked, all warnings will be suppressed.
If checked, enter the temperature value for CO
then be made based on the temperature entered rather than the sample
temperature.
From the drop down list, define the three measurements to display on the right side
of the measurement results screen after the measurement process has completed.
Description
When checked (default), it is required to log on as a registered user to access the
menus. When disabled, all menus are access free and there will be no name recorded
against any actions in the audit trail. To set up valid users, refer to
on page 14.
When checked, the user is logged out automatically when the set delay for inactivity is
reached. The instrument goes on standby. Enter a maximum time of inactivity (in
minutes) for all users.
When checked, every action from a user is recorded in an audit file for traceability.
These actions cover the configuration, calibration, and maintenance options. The audit
file is a rolling buffer recording the last 1000 actions.
normalization. CO
calculations will
2
2
Start up
on page 10. It
User management
English 13