Hach DR 3900 Manual Básico Del Usuario página 20

Ocultar thumbs Ver también para DR 3900:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 45
Error displayed
Definition
Please insert the USB
memory stick.
Please check the
Network setup or FTP
connection and contact
error
the administrator.
Blank value correction
Blank value correction
not possible with
not possible!
LCW919.
File for instrument
Error during update.
update missing.
File for instrument
Error during update.
update is faulty.
It's recommended to
Check of the air values
execute a Full System
failed
Check
Entry invalid!
Password incorrect
The measured
Absorbance > 3.5!
absorbance exceeds 3.5
Fault
Barcode check
Deviation to the stored
number?
data
Please update program
data!
Network setup: DHCP
Error when calling up
client has no connection
the local IP address.
to the DHCP server
English 20
Resolution
Error displayed
Insert a USB memory
Error during default
stick into a USB A port on
gateway setup.
the instrument.
Error during network
drive setup!
Error during subnet
mask setup.
Error copying from USB
Check the USB memory
memory stick.
stick.
Save the update file
again and repeat the
Error in FTP
procedure.
connection.
Switch the instrument of
Fault
and then back on again.
Program not
If the system check is not
accessible.
successful, contact the
Please update program
manufacturer or his
data!
representative.
Forgotten your
Fault
password?
Clean cuvette!
Contact the
manufacturer or his
Fault
representative.
Test program stopped!
Dilute sample and
Please check lamp
measure again
Close the lid.
Error [xx]
Data updating
Fault
Test program stopped!
Please remove the
cuvette
Enter the IP address
Close the lid.
again.
Definition
Resolution
Network setup: default
Try to create the
gateway cannot be set
connection again.
for fixed IP address
Error during network
Check the settings.
setup
Network setup: Subnet
Enter the subnet mask
mask cannot be set for
again.
fixed IP address
Start the procedure again
or contact the
Error during update
manufacturer or his
representative.
Make sure that the
FTP error
instrument is connected
to the network.
Barcode test not present
Data updating
The cuvette is soiled or
Clean the cuvette; allow
there are undissolved
the particles to settle
particles in the cuvette
Check the lamp and
replace it if necessary.
Test program stops when
the instrument is started
Close lid.
Press Start Again.
Remove the cuvette/
sample cell from the cell
Test program stops when
compartment.
the instrument is started
Press OK.