338210r3 Eng.book Page 47 Monday, July 11, 2005 1:28 PM
T
ROUBLESHOOTING
CAUTION
!
If the problem cannot be solved, contact your equipment supplier or ResMed. Do
not attempt to open the S8 Elite case. There are no user serviceable parts inside.
Repairs and internal servicing should only be performed by an authorized service
agent.
Problem
No display.
Treatment pressure
seems low.
Device does not start
when the patient
breathes into the
mask.
Possible Cause
Power is not connected.
Ramp Time is in use.
Air filter is dirty.
Air tubing is kinked or
punctured.
Air tubing is not connected
properly.
Mask and headgear are not
positioned correctly.
Caps are missing from access
port on mask.
Power cord is not connected
properly.
Power outlet may be faulty.
SmartStart not on.
Leak Alert has been enabled;
SmartStart has automatically
been disabled.
Breath is not deep enough to
trigger SmartStart.
Solution
Ensure the power cable is
connected and the power
switch (if available) is on.
Wait for air pressure to
build up.
Replace air filter.
Straighten or replace
tubing.
Check air tubing.
Adjust position of mask and
headgear.
Replace caps.
Connect power cord firmly
at both ends.
Try another power outlet.
Enable SmartStart.
Disable Leak Alert to
enable SmartStart.
Advise patient to take a
deep breath in and out
through the mask.
T
ROUBLESHOOTING
47