R82 Heron Guia Del Usuario página 39

Ocultar thumbs Ver también para Heron:
Tabla de contenido

Publicidad

Test modes during immunity tests:
1.
Battery mode, standby armed/ready.
2.
Charging mode, actuator not running.
EM immunity standards
Electrostatic discharge
IEC61000-4-2
Radiated RF EM fields
IEC 61000-4-3
Proximity
fields
form
wireless
communication
equipment
IEC 61000-4-3
Electrical fast transients /
bursts
IEC 61000-4-4
Surges, Line-to-line
IEC 61000-4-5
Conducted
disturbance
induced by RF fields
IEC 61000-4-6
Rated
power
frequency
magnetic fields
IEC 61000-4-8
Voltage
dips,
interruptions
and
variations
IEC 61000-4-11
support.R82.org
IEC 60601-1-2:2014 test levels
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
3V/m (Professional)
10 V/m (Homecare)
80 MHz – 2.7 GHz
80 % AM at 1 kHz
RF
IEC 60601-1-2:2014 Table 9
± 1 kV, ± 2 kV
5 kHz repetition frequency
± 0,5 kV, ± 1 kV, ± 2 kV
6 Vrms
0.15 MHz - 80 MHz 80% AM
at 1 kHZ
0.15 MHz - 80 MHz 80% AM
at 2 kHZ
30 A/m
50 Hz and 60Hz
short
0 % UT for 0,5 cycle at 0°,
voltage
45°, 90°, 135°, 180°, 225°,
270° and 315° phase angles
0 % UT for 0,5 cycle
0 % UT for 1 cycle
70 % UT for 25 cycles (50 Hz)
70 % UT for 30 cycles (60 Hz)
0 % UT for 250 cycles
Additional test levels
± 6 kV contact
20 V/m
800 MHz – 2.5 GHz
80 % AM at 1 kHz
N/A
N/A
N/A
N/A
N/A
40 % UT for 5 cycles
39