5. Connect power cables to the appropriate power source (90-240 VAC @ 50-60 Hz or 9-36
6. Proceed to Navigating the Menu (Page 26).
FUG1010 IP65 NEMA 4X Quick Start
Operating Instructions, 06/2014, A5E02951527-AC
Pin
TB3
Function
Function
1
AUX. 1 IN
AUX. 3 IN
2
AUX. 1 COM AUX. 3
COM
3
AUX. 2 IN
AUX. 4 IN
4
AUX. 2 COM AUX. 4
COM
Note
If analog input is used for temperature, this will take priority over clamp-on RTD
measurement.
WARNING
Electrical Shock Hazard
Certain parts inside the device carry dangerous high voltage and may result in electric
shock, or serious personal injury.
The transmitter must be grounded and the top cover closed before applying power to
the device.
NOTICE
Power Supply Damage
Improper power connections will damage power supply.
Ensure that all AC or DC power supply connections are properly connected to the
appropriate power source (100-250 VAC @ 50/60 Hz or 9-36 VDC).
VDC). Close transmitter top cover.
TB4
Use
Iin1 Input
Iin1
Common
Iin2 Input
Iin2
Common
3.2 Transmitter Wiring
Description
Behaviour
Analog
4 to 20mA
current
input
referenced
to meter
ground.
Connecting
Load
Wiring
200Ω
305 meters
(1000 ft.)
Max w/o
factory
approval
25