1735
Getting Started
Analyzing the measured values of the recorder function:
These values can be recorded in the instrument, downloaded and evaluated
with Power Log software.
Power
Select Power with the rotary switch.
In this measuring mode you can get the following values for each phase (A, B,
C):
•
Power (P) in W (for each phase and its sum P
•
Reactive power (Q) in var (for each phase and its sum Q
•
Apparent power (S) in VA (for each phase and its sum S
•
Distortion power (D) in VA (for each phase and its sum D
•
Power factor (PF) and the average PF for the three phases.
•
Cos ϕ and the average cosϕ for each of the three phases.
•
Active energy (EP) in kWh.
•
Reactive energy (EQ) in kVAR.
When operating in DELTA connected modes, the Logger will only
display the Ptotal, Qtotal, and the related Power Factor.
20
tot
Note
).
).
tot
).
tot
).
tot