Leveling; Anti-Theft Locking Device - Sartorius Secura 26 Instrucciones De Instalación Y Manejo

Balanzas de laboratorio
Ocultar thumbs Ver también para Secura 26:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 24
Removing the power plug adapter
Press on the ribbed button from above
t
while sliding the power plug adapter
backward.
Push the power plug adapter out of
t
the AC adapter and remove it.
Connecting the AC Adapter
Check the voltage rating on the AC adapter's type plate. Make sure
t
that the voltage rating printed on this unit matches the local sup-
ply voltage at the place of installation.
If the stated supply voltage does not comply with the local supply
t
voltage or there is no suitable AC adapter available: Do not use the
AC adapter. Contact Sartorius Service.
Only use original Sartorius AC adapters.
t
Connect the angle plug (2) to the
t
1
device's power supply socket (operat-
ing voltage connection).
Connect the power cable (1) to the
t
wall outlet (supply voltage) at the
installation location.
2
Warm-up Time
To ensure accurate results are delivered,
0
the balance must warm up for at least
30 minutes after initial connection
to the power supply. Only after this
time will the device have reached the
required operating temperature.
30
When a verified balance of the accuracy class k
for use in legal metrology is connected to the mains
power, it must warm up for at least one hour before
operation.

3.5 Leveling

Leveling the balance compensates for slant or
unevenness at the place of installation. The balance
must be perfectly horizontal to ensure consistent,
reproducible weighing results.
The balance needs to be re-leveled and then adjusted
each time its setup location is changed.
6
Installation and Operating Instructions Manual Secura
Secura Balances:
When "Level" appears on the display, you must level the balance.
®
Quintix
Balances:
Instructions for Use in Legal Metrology
Additional Display Elements:
Metrology line with additional information about minimum load
(Min) and verification scale interval (e).
Interface Specification (Data Output)
Settings "PC – SBI" and "USB RS232 SBI" for the identification
of differentiated scale divisions for use in legal metrology: In the
setting "SBI", there is no automatic identification of display digits
not verified for use in legal metrology. Corresponding measures or
settings must be carried out on the peripheral device.

3.6 Anti-theft Locking Device

®
®
, Quintix
Make sure that the weighing pan is
t
unloaded.
Press "LEVEL".
t
Follow the instructions on the
t
interactive display.
Turn the two feet as shown in the
t
illustration until the air bubble is
centered within the circle of the
level indicator.
− Air bubble at "12 o'clock": turn the
two feet clockwise.
− Air bubble at "3 o'clock": turn the
left foot clockwise and the right
foot counterclockwise.
− Air bubble at "6 o'clock": turn the
two feet counterclockwise.
− Air bubble at "9 o'clock": turn the
left foot counterclockwise and the
right foot clockwise.
Warning symbol to indicate calculated values,
negative gross values or values that are smaller than
the minimum sample quantity (SQmin).
Setup Menu "Settings":
On balances verified for use in legal metrology,
not all functions and settings are available.
A "Kensington" key lock can be
t
installed at the fastening point on
the back of the balance if required.