7.4. Adjusting Setting

IMPORTANT: After powering on Zen-O™, the startup sequence will take approximately 35 seconds.
Specified oxygen level will be reached within 2 minutes of use.
• The device starts working in the previous setting.
Use the mode button
flow mode
• In pulse mode, the device will deliver a pulse of oxygen at the beginning of each of your
inhalation.
• In continuous flow mode, the device will provide a continuous flow of oxygen, but will con-
EN
sume more power and have a shorter battery life.
Setting the mode can be done as follows:
• Pulse mode of operation can be adjusted from 1.0 to 6.0 in 0.5 increments with the
keys.
• Continuous mode of operation can be adjusted from 0.5 to 2.0 in 0.5 increments with the
and
keys.
IMPORTANT: If an air leak is suspected, leaks can be detected with a solution of soap and water ap-
plied to the cannula-concentrator connection point and looking for bubbles.
IMPORTANT: Flow can be verified by putting the oxygen concentrator in continuous mode and plac-
ing the end of the nasal cannula under the surface of a half full cup of water and looking for bubbles.
The current setting and power source (external power or battery; battery icon also shows ap-
proximate level of charge remaining) are shown on the display screen as shown in Fig. 7.
Zen-O™ operating
on battery power.

7.5. Battery Button

The battery button
pushing the button will cycle through all the information.
• First, the gauge information for both batteries (or one battery if only one is installed), will be
shown
• Next, the gauge for just the battery in the first slot
charge cycles on the battery in the first slot
• Finally, the gauge and charge cycles for the battery in the second slot is shown
If there is no battery in one of the slots, then a question mark will be shown instead of the fuel
gauge and number of cycles. After the fifth push of the battery button, the display will alter-
nate back to the main screen showing the current flow setting. It will also automatically exit
the battery status menu and go back to the main flow setting display after approximately 15
seconds of no buttons being pushed.
to alternate between pulse mode
.
Pulse mode
Continuous flow mode
allows you to check the status of the battery or batteries. Repeatedly
.
.
12/185
Fig. 7
.
and continuous
and
Zen-O™ running
on external power.
, then the number of