7 .2 . TURNING ON

• To turn the device on, press the power button
• The concentrator will chirp and the green, yellow, and red LEDs will flash once, while
the screen displays the device name.
Red LED - indicates a warning danger and/or a need for urgent action
Yellow LED - indicates caution or attention required
Green LED - indicates device is on. The green LED will then stay lit.
IMPORTANT: No adjustments can be made until the startup sequence is completed.

7 .3 . CHOOSING A PREFERRED LANGUAGE

• While the device is on, hold down the plus
four seconds until it says "Language:".
• Next cycle through the available languages using the plus
• When the desired language is shown, press the mode button
will change the language and go back to the normal flow screen.

7 .4 . ADJUSTING SETTING

IMPORTANT: After powering on Zen-O™, the startup sequence will take approximately 35
seconds. Specified oxygen level will be reached within 2 minutes of use.
• The device starts working in the previous setting.
Use the mode button
ous flow mode
• In pulse mode, the device will deliver a pulse of oxygen at the beginning of each of your
inhalation.
• Auto Mode: If no inhalation is detected for 60 seconds when in pulse mode, the "Check
Cannula" alarm will be activated and the device will automatically enter Auto-Mode and
continue to deliver oxygen at a rate of 18 breaths per minute. When an inhalation is
detected, the device will clear the "Check Cannula" alarm and exit Auto-Mode.
• In continuous flow mode, the device will provide a continuous flow of oxygen, but will
consume more power and have a shorter battery life.
Setting the mode can be done as follows:
• Pulse mode of operation can be adjusted from 1.0 to 6.0 in 0.5 increments with the
and
buttons.
• Continuous mode of operation can be adjusted from 0.5 to 2.0 in 0.5 increments with
the
and
buttons.
to alternate between pulse mode
.
17/338
.
and mute
buttons together for about
or minus
to select. The device
EN
buttons.
and continu-

Publicidad