Storage
The connector should be protected from
humidity by the protective cap
Place the electrode in the storage protec-
tor
with KCl 3M.
Close the refilling port with the stopper
Life of an electrode
The average life expectancy of a pH elec-
trode is approximately one year depending
on the working conditions.
An electrode used for more than six months
that does not response appropriately after re-
generation, must be replaced by a new one.
If an electrode has too short lifespan,
less than three months, ask us about it.
Probably there is a more suitable elec-
trode for the application.
Specifications
Measuring range
Operating temp.
Reference
element
Diaphragms
Electrolyte
Body material
Min. immersion depth
Diameter
Applications
Biology and clinical areas.
Limitations
Viscous solutions or samples with colloids.
Distilled water.
Warranty
HACH guarantees this electrode against
manufacturing defects.
Validity: 6 months from shipment date.
Limitations:
• Accidental damage.
• Inadequate applications.
• Non-fulfilment of the recommendations.
• Normal wear and tear of the electrode.
QUALITY CERTIFICATE
The attached electrode 52 09 has passed
the quality test and follows the next speci-
fications:
• Asymmetry potential < ± 15 mV.
• Sensitivity, pH 4...7 (at 25 ˚C) > 98%.
• Response time, pH 4...7 < 20 s.
0... 14
0... 80 °C
encapsulated
Ag/AgCl crystals
ceramic
KCl 3M
glass
8 mm
6 mm