OBJ_DOKU-188704-009.fm Seite 29 Donnerstag, 25. März 2021 11:48 11
If, however, you have called up an individual program, the gas
flow for this program is set.
All other operating parameters are not automatically set.
1. Screw the argon gas connection of the handle into the argon
receptacle of the APC unit. Erbe recommends using the ad-
ditional membrane filter, Art. No. 20132-059.
2. Insert the HF connector of the handle into the multifunction-
al receptacle of the APC unit.
3. Make sure that instrument number 4 is displayed on the APC
unit.
4. Only when using an HF electrode: Switch the argon flow to
the APC unit off (argon flow = 0 l/min.). Otherwise, a warn-
ing signal is emitted when activating the handle and an error
message appears.
5. Only select unit settings that do not exceed the maximum
electrical capacity of the product (HF peak voltage/V
The HF voltage that the unit emits depends on the mode
used and the power selected. You can find information about
the technical data of the unit settings in the user manual of
the electrosurgical unit used.
6.5
Conducting a performance test
Before every use conduct a performance test.
The APC-handle can be activated with a finger switch or foot
switch. These instructions describe the function test with finger
switch activation. It is just as good to conduct the function test
with foot switch activation.
1. Press the yellow CUT button.
– For APC 300 + ICC: The respective CUT mode must be ac-
tivated on APC 300 and ICC. If this is not the case, the
handle is damaged.
– For APC 2 + VIO: The CUT mode must be activated on the
VIO. If this is not the case the handle is damaged.
29
max).
p