Processing step
Items Needed
Manual cleaning/
Soft plastic brush/soft disposable cloth
disinfection
(low particle)
50 ml disposable syringe with male
Luer-Lock connection (4 ea.)
Female-to-female luer-lock adapter
(=irrigation adapter)
Compressed air gun (2 bar maximum)
Machine cleaning/
Female-to-female luer-lock adapter
disinfection
(=irrigation adapter)
7.5
Precleaning
Use water for precleaning, a non-fixing disinfectant if necessary.
1. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.
2. Screw the luer-lock adapter onto the argon gas connection
of the handle.
3. Completely fill a disposable syringe with fresh water and
rinse the lumina (at least 5 times).
4. Remove the syringe. Leave the irrigation adapter on the
product for the next processing steps.
7.6
Manual cleaning and disinfection
Use a liquid detergent suitable for the preparation of a dipping
bath.
Use a disinfectant compatible with the detergent suitable for the
preparation of a dipping bath.
32