Three discrete inputs are provided for switch closure inputs or logic level voltage inputs. Make wiring
connections and configure jumper settings to the controller as shown in
Figure
10.
Note:
Figure 9
shows the back of the controller cover and not the inside of the main controller compartment.
Figure 9 Discrete input wiring connections
Discrete inputs
Input 1+
Input 1-
Input 2+
Input 2-
Input 3+
Input 3-
Figure 10 Jumper settings
1 Input 1 configuration
jumpers
2 Input 2 configuration
jumpers
20 English
Table 4 Input connections
Connector position - Switch
input
3
2
6
5
8
7
3 Input 3 configuration
jumpers
4 Jumpers positioned to the
left for switch inputs
Figure
9,
Table 4
and
Connector position - Voltage
input
2
3
5
6
7
8
5 Jumpers positioned to the
right for voltage inputs