"Auto" means automatic measurement. When this function is selected, the instrument
automatically selects an adequate range and pulse width for the measurement. The process of
measurement does not require any intervention by the user.
•
"Auto" is the default setting.
Pulse Width Configuration
The.selection.of.pulse.width.affects.the.dynamic.range.and.resolution.of.the.measurement..With.
a narrow pulse width there will be higher resolution and smaller deadzone; however, the dynamic
range.will.be.decreased..A.wide.pulse.width.will.bring.higher.dynamic.range.and.measure.
comparatively longer distance, but resolution and deadzone will be increased. Therefore, users
should.make.the.choice.between.dynamic.range.and.deadzone..
The.options.for.pulse.width.will.change.according.to.the.distance.range.selected.
Under the parameter configuration menu use p.and.q to highlight "PulseWidth." Press Enter.to.
select a value (Figure 3-11). Press
Use.p.and.q.to.highlight.the.pulse.width..Press.Enter to confirm.
Notes:
•
"Auto" is the default setting.
•
When the range is set to "Auto," the pulse width automatically is set to "Auto."