Descargar Imprimir esta página

Installation - Agilent Technologies TV 301-P Manual De Instrucciones

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 43
All manuals and user guides at all-guides.com

Installation

Do not install or use the controller in an environment exposed to
atmospheric agents (rain, snow, ice), dust, aggressive gases, or in
explosive environments or those with a high fire risk. During
operation, the following environmental conditions must be respected:
temperature: from +5 °C to +40 °C
relative humidity: 0 – 95 % (non-condensing)
The TV 301-P Navigator belongs to the second installation (or overvoltage)
CAUTION!
category as per directive EN 61010-1. Connect the device to a mains line that
satisfy the above category.
The TV 301-P Navigator has Input/Output and serial communication connectors
that must be connected to external circuits in such a way that no electrical part
is accessible.
Be sure that the insulation of the device connected to the TV 301-P Navigator is
adequate even in the case of single fault as per directive EN 61010-1.
For installation of optional accessories, see "Technical Information".
TV 301-P Navigator Controller User Manual / 87-901-001-01 (B)
16
Instructions for Use

Installation

167/224

Publicidad

loading

Este manual también es adecuado para:

969-8994