Technical Information
Profibus message mapping
23
1
°C
24
1
°C
25
1
°C
26
1
-
27
1
Bit field
210/224
All manuals and user guides at all-guides.com
0-100
Controller Power Section Temperature
Read the controller power section temperature.
If > 75 °C the controller goes in fail "controller over-
temperature"
0-100
Controller CPU Section Temperature
Read the controller CPU section temperature.
If > 80 °C the controller goes in fail "controller over-
temperature"
15-45
Pump Body Temperature Limit
Not used
0-6
B0-B3 – Status
0 = Stop
1= Waiting for interlock (interlock connections on
remote I/O connector missing)
2 = Ramp (the pump is increasing the speed)
3 = Auto tuning (the speed has been reduced because
the gas load is higher than the Power Limit)
4 = Braking (the pump is reducing the speed using the
motor)
5 = Normal operation (the pump is at target speed)
6 = Fail (see error code byte Input Buffer Byte-27)
B4-B7 – Reserved
Error Code (bit field)
B7 – Too High Load
(the gas load is too high for the pump)
B6 – Short Circuit
(a short circuit happened between the two of the three
motor phases)
B5 – Over-voltage
(not used)
B4 – Verify Line Select
(the auto voltage selection mismatch the real input
voltage)
B3 – Power Fail
(internal circuitry failure)
B2 – Controller Over-Temperature
("Controller Power Section Temperature" or "Controller
CPU Section Temperature" are over the limit)
TV 301-P Navigator Controller User Manual / 87-901-001-01 (B)