Descargar Imprimir esta página

Agilent Technologies TV 301-P Manual De Instrucciones página 207

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 43
All manuals and user guides at all-guides.com
16
1
Bit field
TV 301-P Navigator Controller User Manual / 87-901-001-01 (B)
B1 - Vent & Purge Remote/Profibus
0 = serial or Profibus (see Out. Buff. byte-0 B3 and B2)
1 = remote ( the valves are driven by the inputs on
remote I/O connector)
B2 - Interlock Type
0 = continuos (the interlock must be present to run the pump)
1 = at start (the interlock must be present only at start
command)
B3 - Setpoint Logic
0 = active close (NO)
1 = active open (NC)
B4 - Heather Jacket Mode (Auto/man)
Specific option, not available on standard models
B5 - Gas Load Type
0 = Ar
1 = N2
B6 - Purge Gas Type
0 = N2
1 = Ar
B7 - Reserved
B0.. B1 - Reserved -
B2.. B3 - Start/Stop/Interlock Mode
Change the logic of Start/stop and Interlock inputs.
Ask to Agilent personnel for this option
B4.. B5 - Set Point Type
0 = frequency
1 = current
2 = time
3 = status (normal operation)
B6.. B7 - Analog Output Type
Set the analog output on remote I/O connector
proportional to:
0 = frequency
1 = power
Technical Information
Profibus message mapping
17
146
110
104
-
157
-
114
151
101
111
207/224

Publicidad

loading

Este manual también es adecuado para:

969-8994