Table 12
Input Buffer (15 words)
Offset
Size
Unit res
byte
byte
0
2
mA
2
2
W
4
2
W
6
2
Hz
8
2
Hz
10
2
-
12
2
-
14
2
-
16
2
Min
18
1
Hours
20
1
V
21
1
°C
22
1
°C
TV 301-P Navigator Controller User Manual / 87-901-001-01 (B)
Set the high rotational speed, default = 530
500-530
Low Speed Value
Set the low rotational speed, default = 500
Range
Description
0-10000
Current
Motor current consumption
0-750
Power
Motor power adsorption
0-750
Power Limit Applied
Maximum allowable power
Driving Frequency
0-530
Read the rotational speed of the pump
0-530
Target Frequency
Read the target rotational speed
-
Flow Meter Alarm Time
Specific option, not available on standard models
Flow Meter Alarm Event
-
Specific option, not available on standard models
0-65536
Cycle Number
Number of cycles (start and stop) done by the pump
0-65536
Last Cycle Time
How long the last cycle lasted
0-65536
Pump Life
Total pump running time
0-100
Voltage
Voltage provided to the motor
0-70
Pump Bearing Temperature
Read the pump upper bearing temperature.
If >60°C the controller goes in fail and the pump is stopped
0-45
Pump Body Temperature
Read the pump body temperature close to the water circuit.
If >42°C the controller goes in fail and the pump is stopped
Technical Information
Profibus message mapping
17
117
Win
serial
200
201
155
203
120
213
215
301
300
302
201
204
216
209/224