Cover unused module slots with a Slotcover.
4.1.2
Technical data of the Base Unit
Network connection termi-
nal:
Operating voltage:
Current consumption:
Power loss:
SFR output:
SSR output:
Pre-alarm relay output:
Main alarm relay output:
Monitoring functions
Terminal Blocks:
Terminal type:
Cable core cross-section:
WARNING
Risk of electric shock!
If live cables with a voltage of max. 240 V AC are connected to the Docking Station:
► To ensure electrical safety, the Docking Station may only be used in the closed
housing (Base Unit with Display Unit) or in a cabinet.
► Live terminals are accessible after opening the housing or cabinet. If necessary,
turn off the mains power supply when working on the housing or cabinet
installation.
► The connection lines used must at least have basic insulation.
|
Instructions for use
REGARD 3000
Wire cross-section min. 18 AWG
100 – 120 V AC / 200 - 240 V AC / 50 / 60 Hz
3 A max. (type 1 A)
(depending on the number of installed modules and
connected transmitters)
Max. 25 W (10 W (power supply) + 15 W (Docking
Station))
min. 5 V, 10 mA, max. 30 V, 2 A switching capacity;
Protect the output overload
min. 5 V, 10 mA, max. 30 V, 2 A switching capacity;
Protect the output overload
min. 5 V, 10 mA, max. 30 V, 2 A switching capacity;
Protect the output overload
min. 5 V, 10 mA, max. 30 V, 2 A switching capacity;
Protect the output overload
– Input voltage monitoring 18-30 V
The SFR is activated for voltage underrun; the
power supply is shut down in case of voltage over-
run and the SFR is activated, then a restart is
required.
– Total current monitoring
Shut-down at >5.5 A (max. 6 A), then a restart is
required
– Temperature monitoring
Shut-down of the Docking Station and activation of
the SFR, then a restart is required
plug-in terminal
Min. 18 AWG
|
System components
109
en