Transport; Disposal; Technical Data; General Data - Dräger Polytron 6100 EC WL Instrucciones De Uso

Ocultar thumbs Ver también para Polytron 6100 EC WL:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 96
en
|

Transport

15
Transport
The gas detector contains lithium batteries. During transportation, especially
when transporting the gas detector by air, observe the relevant safety
regulations and marking for lithium batteries.
16

Disposal

This product must not be disposed of as municipal waste. It is therefore
marked with the adjacent symbol. Dräger will take back this product for
free. For information please contact the national marketing organiza-
tions or Dräger.
Batteries must not be disposed of as household waste. They are there-
fore marked with the adjacent symbol. Dispose of batteries and
rechargeable batteries at battery collection points and in accordance
with current regulations.
Disposal of electrochemical sensors
WARNING
Danger of explosions and risk of chemical burns!
Sensor fluids may leak out and can cause acid burns.
► Do not dispose sensors in fire.
► Do not open with force.
17

Technical data

17.1

General data

Housing material
Stainless steel
Housing protection class
IP 66/IP 67
62
Instructions for use
|
Dräger Polytron
Weight of gas detector (without
approx. 3.2 kg (7.0 lb)
battery pack)
Weight of battery pack
approx. 0.6 kg (1.3 lb)
For information about the dimensions, see appendix "Dimensions".
17.2

Power supply

Wired power supply
Power supply unit
The power supply must comply with the
National Electric Code (NEC) Class 2 or
Protection Class II specification.
Cable type
2-core cable
Conductor
24–12 AWG / 0.2–2.5 mm²
Cable entry
Ex e certified, M 25 x 1.5
Operating voltage
14–30 V
Inrush current
0.5 A for 3 ms at 30 V
resistance;
0.7 A for 3 ms at 30 V
resistance
Power consumption
0.5 W
Operational parameters for
U
power supply with safety bar-
1)
rier
1)
For long cables, the cable resistance can cause voltage drops of the input voltage.
®
6100 EC WL, Dräger Polytron
, 10 Ω conductor
, 1 Ω conductor
= 30 V, P
= 1.2 W, C
= 20 nF, L
= 0 mH
i
i
i
i
®
Repeater WirelessHART
®

Publicidad

Tabla de contenido
loading

Tabla de contenido