1.4. Units
Sensor measurement units.
Options

1.5. Operating Mode

space
level
Operation
Description
NO
Measurement and associated loop current are not
SERVICE
updated, and the device defaults to Fail-safe mode
LEVEL
Distance to material surface
SPACE
Distance to material surface
DISTANCE
Distance to material surface
a)
For more details on Fail-safe Mode refer to the full manual.
1.6. Low Calibration Point
Distance from sensor reference point to Low Calibration Point: usually process empty level.
Range: 0.00 to 20.00 m
Values
1.7. High Calibration Point
Distance from sensor reference point to High Calibration Point: usually process full level.
Range: 0.00 to 20.00 m
Values
1.8. Apply? (Apply changes)
In order to save the Quick Start settings it is necessary to select Yes to apply changes.
Options
Press Mode
to return to Measurement mode. SITRANS LR200 is now ready to operate.
Page EN-18
SITRANS LR200 (HART) – QUICK START MANUAL
M, CM, MM, FT, IN
distance
high cal.
point
YES, NO, DONE (Display shows DONE when Quick Start is successfully
completed.)
sensor
reference
point
low cal.
point
Reference point
a)
.
Low Calibration Point
High Calibration Point
Sensor reference point
A5E31993614