Flow chart:
Cover the hole on the inside of the socket with a small piece of paper and tape that is not
airpermeable (such as PVC tape). Cycle the pump until a vacuum of 508-847mbar (15 – 25 inHg)
is reached. Check if vacuum level is maintained when the pump is not being cycled.
Vacuum maintained:
Zone 2 and 3 are OK
Examine Zone 1 (socket and sleeve) for any
leaks
With T-Connector removed, attach the Vacuum Gauge directly to the Intake Valve (Fig. 1, item 5)
and cycle the pump until 508-847 mbar (15 – 25 inHg) is reached. Check if vacuum level is
Vacuum maintained:
Zone 3 is OK
Problem is in the Socket Connector or in its
bonding with the socket. Examine for leaks or
damage.
Vacuum maintained:
Replace the Valves
4.7 Cosmetic finish of the prosthesis
The Harmony P3 is delivered with two different exhaust valve attachments:
• a short piece of tubing with installed muffler and Double Barb Connector pre-installed at the
exhaust valve
• a long piece of tubing with an exhaust flange
If a cosmetic cover of the prosthesis is desired use a cover with a 34 mm centre hole.
NOTICE
Function loss as a result of improper installed cosmetic cover. If using a 6R8 Foam Cover,
it is important that the cover is not compressed along the length of the foam, as this can affect
the function of the Harmony pump.
Once the Foam Cover is in place, determine location of exhaust flange and drill or cut a 6 mm
hole. Ottobock recommends placing the exhaust flange medially at the ankle, just above the top
of a patient's shoe. Next, insert tubing (Fig. 12, Pos. 2) and flange (Fig. 12, Pos. 3) into the hole
of cosmetic cover and connect tubing to Double Barb Connector (Fig. 12, Pos. 1). Make sure that
­armonyy3 44147
All manuals and user guides at all-guides.com
maintained when the pump is not being cycled.
Exchange Valves and check again
Replace the Functional Ring
Vacuum not maintained:
Problem exists in Zone 2 or Zone 3
Vacuum not maintained:
Problem exists in the Valves or
Functional Ring
Vacuum not maintained:
Problem exists in the Functional Ring
Ottobock | 19