10. Click Start-depending on the size of your thumb drive these may take a while. When the
format is complete you will receive a pop-up indicating it has formatted successfully
11. Copy "update_user.logo.bin" file onto the formatted thumbdrive
12. It is very important that this file and only this file is present on the USB thumb drive
13. With the audiometer turned off insert the thumb drive into any available USB port
14. Turn on the AD629 and push the Temp/Setup Button from the Tone test screen
15. Enter "Common Settings" using the Setup/Tests Button
16. For the question "Do you want to install" press the "Yes" button
17. After installation is completed, press the "Back" button to get to the testing screen
AD629 - Instructions for Use - EN
Page 34