O O p p e e r r a a t t i i o o n n
A A p p p p l l y y i i n n g g o o r r r r e e l l e e a a s s i i n n g g t t h h e e b b r r a a k k e e s s
W W A A R R N N I I N N G G - - Always apply the brakes when a patient is getting on or off the product or when the product is not moving.
Injury may result if the product moves while a patient is getting on or off the product.
To apply the brakes, push down on the brake (A) (Figure 2) side of pedal ( Product illustration (page 6)).
To release the brakes, push down on the steer (C) (Figure 2) side of pedal ( Product illustration (page 6)).
A
Brake
N N o o t t e e - - Clean the bottom of the brake pads regularly to prevent wax or floor remnant buildup.
R R a a i i s s i i n n g g o o r r l l o o w w e e r r i i n n g g t t h h e e l l i i t t t t e e r r
W W A A R R N N I I N N G G
• Always place, remove, or transfer the patient from the center (seat) section of the product unless the product is in the full
chair position. Product instability or tip may occur.
• Always put the product in the chair position with the siderails up and latched when you leave a patient unattended on the
product. Do not leave the product in the horizontal (flat) position to avoid a patient fall.
• Do not sit on either end of the product. The product may tip.
• Always keep the product in the chair position when not in use.
C C A A U U T T I I O O N N
• Always remove any devices that may be in the way before you raise or lower the litter.
• Do not raise the product (hydraulics on base) with a patient lift under the product.
To raise the litter, pump the pump pedal (A) repeatedly until you achieve the desired height (Figure 3).
To lower the entire litter, press on both pedal B and pedal C at the same time.
To lower the head end of the litter, depress pedal (B).
To lower the foot end of the litter, depress pedal (C).
EN
8
A
B
F F i i g g u u r r e e 2 2 – – B B r r a a k k e e / / s s t t e e e e r r f f u u n n c c t t i i o o n n s s
B
Neutral
C
C
Steer