The C C h h a a p p e e r r o o n n e e option allows you to set zone controls to alert an operator when a patient may be attempting to exit the
product. The C C h h a a p p e e r r o o n n e e option is available for products with the electric litter option and the scale system option. The
scale system option with C C h h a a p p e e r r o o n n e e option has a battery backup option. The standby icon indicates when the product is
using the battery backup.
Raise the foot end push handles before you attempt to operate the scale system option.
Use the scale system control panel to operate the scale system option ( Scale icon/button identification (page 29)).
Do not touch the product while you weigh or zero the scale system option.
The patient must remain still while the system is weighing. If the patient is moving, the scale system will try for 20 seconds
to get a stable weight or zero value before displaying the error message (
The error (
) appears in the display if there is a loose connection or if the scale is not operating. If you attempt to
operate the scale again and the error appears, contact Stryker Technical Support.
N N o o t t e e - - To meet the scale system accuracy claim, the patient surface must be in the flat position (Fowler backrest and Gatch
down) and the product cannot exceed 5 degrees of Trendelenburg/Reverse Trendelenburg.
S S c c a a l l e e i i c c o o n n / / b b u u t t t t o o n n i i d d e e n n t t i i f f i i c c a a t t i i o o n n
I I t t e e m m
I I c c o o n n / / b b u u t t t t o o n n
A
B
C
D
E
F
G
W W e e i i g g h h i i n n g g a a p p a a t t i i e e n n t t
C C A A U U T T I I O O N N - - Do not use the scale system reading as a reference for medical treatment.
1115-409-005 Rev AB.0
N N a a m m e e
D D e e s s c c r r i i p p t t i i o o n n
Stand by icon
Illuminates green when you plug the power cord into a wall outlet.
Illuminates amber when you unplug the power cord from the wall outlet.
Battery backup functionality will turn on.
Zone 1 icon
The patient can move freely, but the alarm sounds when the patient
attempts to leave the product.
Zone 2 icon
The patient can move with limited movement. The alarm sounds when
the patient approaches the siderails or moves toward the foot end.
Arms and disarms the C C h h a a p p e e r r o o n n e e option (stretcher exit). See Arming
Arm/Disarm button
or disarming the Chaperone option (page 30).
Weigh button
Weighs the patient. See Weighing a patient (page 29).
lb/kg button
Changes the displayed unit of measure. See Locking the scale unit of
measure (page 30).
Zeros the scale settings. See Locking the scale unit of measure (page
Zero button
30).
).