Calibration Checks - Testing Instructions; (For Trained Specialist Staff Only) - boso medicus exclusive Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

medicus_exclusive_5spr_1402sd.qxd

Calibration checks - Testing instructions

(for trained specialist staff only)

Calibration checks – every 2 years at the
latest – must be performed by the follow-
ing institutions or persons:
– Manufacturer
– Trade measurements authorities
– Persons who fulfil the requirements of
the legislation governing the operation
of medical devices.
A) Function testing
Function testing of the monitor can
only be carried out on a person or with
a suitable simulator.
B) Testing of pressure circuit integrity
and deviation of pressure display
72
All manuals and user guides at all-guides.com
10.02.2014
14:47 Uhr
a) When the pressure in measurement
b) A pump ball must also be introduced
1.) Remove batteries.
2.) Hold START button pressed down and
Seite 72
Note:
mode increases above 299 mmHg, the
quick release valve is triggered and the
pressure circuit opens. When the press-
ure in test mode increases above 320
mmHg, the measured value display
flashes.
into the pressure circuit for the tests.
Testing
insert batteries.

Publicidad

Tabla de contenido
loading

Tabla de contenido