OHAUS Valor 1000 Manual De Instrucciones página 8

Ocultar thumbs Ver también para Valor 1000:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 11
EN-6
RS232 Input
The scale will respond to various commands sent via the interface adapter.
Terminate
Commands
with
[CRLF]
IP
[CRLF]
P
CP
[CRLF]
SP
[CRLF]
xS
[CRLF]
[CRLF]
xP
Z
[CRLF]
T
[CRLF]
[CRLF]
PU
xU
[CRLF]
[CRLF]
PV
xFMT
[CRLF]
[CRLF]
\EscR
W
W
[CR]
$
\DC1
\DC2
\ENQ
Description
Immediate Print of displayed weight (stable or
unstable).
Print displayed weight (stable or unstable) with
pre-set format, e.g. "0FMT"
Continuous Print.
Print on Stability only once.
0S: Set Stable menu to "off"
1S: Set Stable menu to "on"
Interval Print x = Print Interval (1-3600 sec), 0P
turns auto print OFF.
Same as pressing Zero Key.
Same as pressing Tare Key.
Print current unit: g, kg, lb, oz
Set scale to unit x: 1=g, 2=kg, 3=lb, 4=oz, 5=lb:oz
Version: print name, software revision and LFT
ON (if LFT is set ON).
0 = (default, Scout); 1 = print format type 1 (NV);
2 = print format type 2 (RE)
Global reset to reset all menu settings to the
original factory defaults
Output same as "P" command
Output same as "P" command
Output same as "P" command
Output same as "P" command
Output same as "P" command
Return \ACK

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Rs232

Tabla de contenido