5.6

Connect input and output signals

Input and output signals can be connected to the terminal strip on the printed circuit board.
Clamp
Function
ID1
Remote start/stop
UA1
Oxygen concentration
UA2
Outlet pressure
K5
Start stop signal to external booster
K6
General alarm (nc/no)
K7
General alarm (nc/no) (=K6)
K8
Spare
The circuits connected to the relay contacts must be supplied by a fused or current limited
power supply. All cables must be double insulated.
K3.1.142M Manual NitroFlow Lab
Input/output signals
Digital input
Nominal input current: 10 mA
Voltage: internal power supply
Analogue output
Input impedance to reach
0 mA – 20 mA: 200 Ohm
Volt Free Relay Contacts
Maximum Switch voltage: 48V AC/DC
Switch current: 1A AC/DC
- 15 -