Parker NitroFlow Lab Guia Del Usuario página 20

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

A
Button
Selection
Active
No
Active
Yes
Active
Auto reset
Stop
Yes
Stop
No
Level
0-16% O
100 – 84% N
Level
0-13 BAR*
0-188.5 PSI*
Level
0-10 BAR*
0-145 PSI*
Delay
0-300 sec
*see also local settings
ATTENTION:
It is impossible to set O
O
high has been set. Therefore, first set O
2
K3.1.142M Manual NitroFlow Lab
Result
Alarm function for this parameter is not active
Alarm function for this parameter is active; alarm messages must
be reset manually
Alarm function for this parameter is active; When alarm level is
not exceeded any longer before manual reset, the alarm will reset
itself
Generator will switch off in case alarm level is exceeded
An alarm signal will be given but generator will continue to run in
case alarm level is exceeded
For screen 1/6 and 2/6: this is the oxygen- or nitrogen level* at
2
which the alarm is set.
2
For screen 3/6 and 4/6.
This is the pressure level at which the alarm is set
For screen 5/6 and 6/6.
This is the pressure level at which the alarm will appear.
Delay time in seconds between the moment that the alarm level
has been exceeded and signaling; this feature prevents false
alarms in case of short spikes
-menu
low at a higher level than O
2
high. The setting of O
2
high level before setting the O
2
low is limited once
2
low level.
2
- 19 -

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido