Data Logging Menu - Parker NitroFlow Lab Guia Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Replace O2 cell
6.3.7

Data logging menu

Access:
Press data logging menu button
(refer to § 6.3)
Function:
Read the logged (saved) data
Alarms as well as status of the sensors are saved on the SD-card. The time between the logging (saving) of this
data to the memory card can be chosen in the data-logging menu.
Button
Interval
A
B
Button
A
B
C
CAUTION:
Please check the alarm, p-switch, options and settings before you restart the unit.
The unit cannot run without the SD-card. This will generate an alarm (SD-card failed).
K3.1.142M Manual NitroFlow Lab
When an O
cell has been replaced during maintenance, this button can be
2
touched and a new date to replace the O
confirmation. In maintenance screen 1/5 the O
from date of changing.
Selection
Result
30-3600
Time in seconds between the logging (saving) of alarm
data
Shows all the alarms that have been saved on the memory
drive (see below)
Explanation
Date and time of alarm incident
Alarm description
The number of logged alarms
cell is set. System asks for
2
2
in settings screen
lifetime should read 3 year ahead
C
- 24 -